Product Number: 6849400
Sodium monitoring for high purity waters
The 9245 Sodium Analyzer is a single channel online analyzer with a detection limit of 0.01 ppb and a measuring range of 0-10,000 ppb. To maintain optimum response time, even in systems of continuous low sodium concentration, the analyzer automatically reactivates the electrode using non-hazardous chemicals. This sodium analyzer is easy to install; it requires only power and sample connections. Four sets of isolated analog outputs can be configured in 4-20 mA outputs and one reporting electrode live signal including calibration and grab sample.
To maintain optimum response time, even in systems of continuous low sodium concentration, the analyzer automatically reactivates the electrode using non-hazardous chemicals.
The 9245 Sodium Analyzer installs easily; it requires only power and sample connections. Clear step-by-step instructions are given by the analyzer to simplify operations such as instrument start-up, long-term stand-by, and reagent replenishment. Low ppb sodium solutions for calibration can be difficult to prepare so the 9245 analyzer is equipped with a fully automated calibration system that uses 10 ppm samples and follows a multiple step calibration cycle to accurately calibrate the analyzer. Normal maintenance consists of replenishment of the calibration solution reagents and electrolyte, typically every 100 days, and annual replacement of the reagent tube.
The analyzer is fitted with a quick-loop valve, a constant head vessel, and minimum sample flow detection to allow for variations in sample pressure, flow, and distance from the analyzer to the fresh-sample point. Grab sample can be obtained for manual calibration, accuracy verification, and measurement of one-off process samples after which, the unit returns automatically to on-line monitoring.
With a detection limit of 0.01 ppb and a range of 0 to 10,000 ppb, the 9245 Sodium Analyzer is ideally suited for monitoring sodium in demineralized water, boiler feed, condensate, and all parts of the steam/water cycle.
Four sets of isolated analog outputs can be configured in 4-20 mA with three outputs assigned to sample concentration or temperature readings and the fourth reporting electrode live signal, including calibration and grab sample. Additional digital communication is available with JBUS/MODBUS or Profibus DP.
Two programmable alarm relays may be assigned to concentration limits, delay, hysteresis, normal relay status, minimum sample flow rate detection, or operation of the measurement cycle for a channel. Two additional programmable relays may be used for warning messages (reagent level low, small calibration deviation, and system alarms (no reagents, no sample, no calibration, no power supply).