CA610 Fluoride Analyzer with Reagents

CA610 Fluoride Analyzer with Reagents

Product Number: 5740001

Ships within 3-5 weeks
Price: Contact Us
Talk to an Expert

Ion-selective electrode with patented, replaceable tip. Automatic calibration. Temperature-controlled wall-jet flow cell. Rugged, lightweight enclosure. Ultra-low reagent consumption. Virtually immune to interferences.

Applications:

  • Municipal Water

Benefits

Reliable, Full-time Fluoride Monitoring

The CA610 Fluoride Analyzer uses advanced ion-selective electrode (ISE) technology for continuous monitoring of fluoride concentration in drinking water effluent. Accuracy is ensured with precise control of temperature, ionic strength, and pH. Interferences that can bias measurements are virtually eliminated with this electrode.

Patented Electrode Design with Replaceable Tip

The proven ISE technology of the electrode is enhanced by a patented* design and manufacturing process. The lanthanum crystal at the tip of the probe is molded into a one-piece, monolithic tip that is threaded for easy removal. This prevents errors and premature failures. Electrodes are equipped with BNC connectors so that they can be independently checked with laboratory electrochemical meters.

Cost Effective Operation

The electrode of the CA610 analyzer is made with a molded fluoride crystal tip. The working life of the electrode is approximately one year. The specially designed tip of the electrode should be replaced every six months. Reagent consumption is very low and maintenance is simple.

Method of Analysis

The ISE system of the CA610 analyzer includes a reference pH electrode and a working electrode. Three reagents are used: Total Ionic Strength Adjustment Buffer (TISAB) solution and two fluoride standards. The TISAB is used for the following purposes. The Ionic strength adjustment lets the electrode respond to concentration instead of activity. The pH control ensures fluoride is present in its ionic state. The Weak complexes formed by certain metals (for example, aluminum) are broken up by TISAB and chelates potential interferents for accurate fluoride measurement.

Automatic Calibration

Two calibration standards are fully enclosed in the CA610 analyzer's case. The instrument performs a 2-point calibration at 0.5 and 5.0 mg/L fluoride at user-selected time intervals. Automatic calibration intervals can be set for 1 day to 1 month. Readings and calculations of the most recent calibration is stored in the instrument.

Loading